Skip to content
Recall Observatory FDA recall evidence

Device product

Laksell Gamma Knife Perfexion

Z-0152-2015

October 06, 2014

Class II

Product summary

Firm
Elekta, Inc.
Event
Event 69425
Status
Terminated
Classification
Class II
Quantity
181
Official record key
device-enforcement:Z-0152-2015

Official wording

Reason: Elekta has become aware that the latches may be locked before they have been fully turned resulting in a poorly locked frame.

Code information: Serial No. 6052, 6144, 6088, 6122, 6158, 6005, 6028, 6097, 6153, 6064, 6081, 6103, 6120, 6143, 6145, 6213, 6215, 6076, 6070, 6136, 6002, 6090, 6131, 6063, 6160, 6077, 6018, 6031, 6040, 6105, 6121, 6013, 6026, 6038, 6050, 6089, 6042, 6055, 6058, 6071, 6080, 6087, 6091, 6096, 9098, 6106, 6108, 6109, 6116, 6118, 6125, 6132, 6133, 6142, 6148, 6149, 6161, 6162, 6175, 6181, 6183, 6198, 6205, 6206, 6163, 6169, 6033, 6041, 6046, 6068, 6113, 6019, 6067, 6110, 6104, 6043, 6044, 6062, 6085, 6086, 6101, 6111, 6123, 6126, 6157, 6177, 6178, 6024, 6069, 6094, 6007, 6075, 6182, 6065, 6159, 6199, 6003, 6017, 6027, 6039, 6112, 6146, 6194, 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, 6030, 6032, 6034, 6035, 6036, 6037, 6045, 6047, 6048, 6053, 6054, 6056, 6057, 6060, 6061, 6066, 6072, 6073, 6074, 6078, 6079, 6082, 6083, 6084, 6092, 6093, 6095, 6099, 6100, 6102, 6107, 6114, 6115, 6117, 6119, 6124, 6127, 6128, 6129, 6130, 6134, 6135, 6139, 6141, 6150, 6151, 6165, 6167, 6171, 6172, 6173, 6176, 6179, 6180, 6184, 6185, 6191, 6192, 6200, 6203, 6207, 6211

Distribution pattern: Worldwide Distribution - USA (nationwide) including the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OR, PA, RI, SC, TX, TN, UT, VA, WA and WV, and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Mexico, Morocco, Netherland, Norway, Poland, Portugal, Russia, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Elekta has become aware that the latches may be locked before they have been fully turned resulting in a poorly locked frame.