Device product
Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry Catalog Number:1360 Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes
Z-0635-2013
Product summary
- Event
- Event 63809
- Status
- Terminated
- Classification
- Class III
- Quantity
- 7865
- Official record key
device-enforcement:Z-0635-2013
Official wording
Reason: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).
Code information: Carton/Bottle Lot Number: C9093061B P9093061B C9094081B P9094081B C0097091B P0097091B C0100101C P0100101C C0103042F P0103042F C0106062B P0106062B C91 01022D P91 01022D CP9105052B P9105052B C9108092A P9108092A
Distribution pattern: Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
Derived failure modes
-
Unknown
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).