Device product
VITROS Chemistry Products CRP Slides (Reagent), REF/Catalog Number 192 6740 and 809 7990, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products CRP Slides quantitatively measure C-reactive protein (CRP) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Z-1615-2013
Product summary
- Event
- Event 65247
- Status
- Terminated
- Classification
- Class III
- Quantity
- Domestic: 4,257; Foreign: 923
- Official record key
device-enforcement:Z-1615-2013
Official wording
Reason: While performing interval testing Ortho Clinical Diagnostics (OCD) determined that an excess amount of surfactant was unintentionally added to VITROS Chemistry Products CRP Slides during the manufacturing process.
Code information: GEN 3773, Coating 0766, Lot XXXX (exp. 11/01/2013); GEN 3784, Coating 0767, Lot XXXX (exp. 11/01/2013); and GEN 3785, Coating 0768, Lot XXXX (exp. 10/01/2013 and 11/01/2013). The term GEN and Coating refer to specific segments of the reagent lot number.
Distribution pattern: Worldwide Disribution - USA Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela.
Derived failure modes
-
Unknown
While performing interval testing Ortho Clinical Diagnostics (OCD) determined that an excess amount of surfactant was unintentionally added to VITROS Chemistry Products CRP Slides during the manufacturing process.