Skip to content
Recall Observatory FDA recall evidence

Device product

Cloth electrodes with Silver-Carbon construction: P/N 202-9231, Cat No. 400-899, 2" x 2" Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9241, Cat No. 400-895, 2" Round Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9381, Cat No. 400-889, 2" x 3.5" Electrode, Cloth AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9501, Cat No. 202-9501, 1.25" Round Electrode, Cloth AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 400-8991, Cat Nos.: 400-877T and 400-899T, 2" x 2.25" Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.

Z-0816-2013

November 30, 2009

Class II

Product summary

Firm
NAImco Inc dba Richmar Inc
Event
Event 63927
Status
Terminated
Classification
Class II
Quantity
25860 electrodes
Official record key
device-enforcement:Z-0816-2013

Official wording

Reason: Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.

Code information: All Serial Numbers distributed between 11/01/2009 and 01/31/2010

Distribution pattern: Distributed in the states of GA and TN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.