Device product
Dimension Troponin 1 (TNI) Flex reagent cartridge
Z-2606-2017
Product summary
- Event
- Event 76915
- Status
- Terminated
- Classification
- Class II
- Quantity
- 70,428 units
- Official record key
device-enforcement:Z-2606-2017
Official wording
Reason: The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
Code information: Device Listing Number # D056469 Lot # EB7091, EC7104, EA7125, ED7193, EA7161, EB7173, EC7229, ED7236 EA7250, ED7271, EB7315, EB7342 "
Distribution pattern: Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa
Derived failure modes
-
Unknown
The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.