Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.
Z-1714-2014
Product summary
- Event
- Event 68363
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2
- Official record key
device-enforcement:Z-1714-2014
Official wording
Reason: Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop.
Code information: Serial Numbers: ULD1412014, UUC13Z2011, A4A113Z2001.
Distribution pattern: Nationwide Distribution including CA and IN
Derived failure modes
-
Unknown
Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop.