Device product
HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics
Z-0741-2013
Product summary
- Event
- Event 64117
- Status
- Terminated
- Classification
- Class I
- Quantity
- 97 total ventilators; 69 of part number 161005 and 28 of part number 161006
- Official record key
device-enforcement:Z-0741-2013
Official wording
Reason: Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.
Code information: Software Versions 1.1.2 and below; Part numbers 161005, 161006.
Distribution pattern: US Nationwide Distribution including the country of Switzerland.
Derived failure modes
-
Unknown
Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.