Device product

Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.

Z-1184-2021

January 13, 2021

Class II

Product summary

Firm
Medical Components, Inc dba MedComp
Event
Event 87265
Status
Ongoing
Classification
Class II
Quantity
96 pieces total (updated 3/11/2021 additional 72 pieces)
Official record key
device-enforcement:Z-1184-2021

Official wording

Reason: The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.

Code information: Catalog Number: MRDP50AXS, Lot Number: MPDX180, UDI Number: 884908149753 (updated 3/11/2021 added lot number: MPCY660)

Distribution pattern: US Nationwide distribution in the states of CA, DC, GA, IL, IN, MA, MI, NJ, NM, NY, OH, PA, TX. Updated 3/11/2021

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.