Device product
Abbott Acclaim Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide variety of standard administration sets and fluid containers.
Z-1482-2014
Product summary
- Firm
- Hospira Inc.
- Event
- Event 67591
- Status
- Terminated
- Classification
- Class I
- Quantity
- 13,218 infusion pumps total
- Official record key
device-enforcement:Z-1482-2014
Official wording
Reason: Hospira has received customer reports of broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.
Code information: List Number: 12032; Serial Numbers: 12322921, 12323859, 12322780, 12322801, 12322810, 12322812, 12322878, 12322880, 12322887, 12322896, 12322903, 12322904, 12322911, 12322917, 12322936, 12322939, 12323867, 12322916, 12323056, 12321139, 12324767, 12324854, 12324802, 12322873, 12324375, 1231453, 1231610.
Distribution pattern: Nationwide Distribution -- AK, AL, AR, AZ, CA, CO, CR, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
Derived failure modes
-
Unknown
Hospira has received customer reports of broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.