Device product

Ortho Clinical Diagnostics VITROS¿ Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS¿ Vapor Adsorption Cartridges are used to remove vapor contaminants that may potentially interfere with Signal Reagent used with MicroWell Assays on the VITROS¿ 3600, VITROS¿ 5600, and VITROS¿ ECi/ECiQ Systems.

Z-0755-2013

December 13, 2012

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 63934
Status
Terminated
Classification
Class II
Quantity
Domestic: 15,891; Foreign: 5237
Official record key
device-enforcement:Z-0755-2013

Official wording

Reason: Some VITROS¿ Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently installed on the VITROS¿ System instead of the intended VITROS¿ Vapor Adsorption Cartridge which may bias results.

Code information: CAT No. 680 0100; the cartridges do not have lot numbers.

Distribution pattern: Worldwide Distribution, including Nationwide (USA) and the countries of AU, GM, FR, IT, SP, BR, BM, AR, CA, JA, MX, UK, CH, CL, CO, IN, EC, PE, PY, PA, PT, NZ, SG, VE and UY.

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    Particulate