Device product
SERF Novae Dual Mobility Cup Insert 47/28- ApeX-LNK Poly Liner. Intended for use with the Apex Modular Acetabular Cup for total hip replacement procedures. Product Code: H6-24728
Z-0921-2021
Product summary
- Event
- Event 87062
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3 units
- Official record key
device-enforcement:Z-0921-2021
Official wording
Reason: Acetabular Liners have a higher than acceptable oxidation limit, potential for patients to experience a higher wear rate for acetabular liners and require a revision surgery
Code information: Lot Number: 28624
Distribution pattern: AL, ID, MD, NY, OK
Derived failure modes
-
Unknown
Acetabular Liners have a higher than acceptable oxidation limit, potential for patients to experience a higher wear rate for acetabular liners and require a revision surgery