ADVIA Chemistry Systems Drug Calibrator I; For in vitro diagnostic use in the calibration of Phenobarbital_2 (PHNB_2), Phenytoin_2 (PHNY_2), and Theophylline_2 (THEO_2) methods on the ADVIA Chemistry systems. System Information: (For Information Only) The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use. The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use.

Z-1321-2015

January 19, 2015

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 70410
Status
Terminated
Classification
Class II
Quantity
657
Official record key
device-enforcement:Z-1321-2015

Official wording

Reason: Incorrect calibrator lot values were assigned for ADVIA Chemistry Systems Drug Calibrator I, lot 3JD018, for the ADVIA Chemistry Systems Phenytoin 2 (PHNY_2) and Phenobarbital 2 (PHNB_2) assays which was the result of calibrator value miscalculation.

Code information: Material Number 10376770, Lot Number 3JD018, Expiration: April 2015

Distribution pattern: Worldwide Distribution - US (nationwide) and Internationally to Canada and Mexico.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Incorrect calibrator lot values were assigned for ADVIA Chemistry Systems Drug Calibrator I, lot 3JD018, for the ADVIA Chemistry Systems Phenytoin 2 (PHNY_2) and Phenobarbital 2 (PHNB_2) assays which was the result of calibrator value miscalculation.