Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.

Z-0197-2015

October 24, 2014

Class II

Product summary

Firm
Integra LifeSciences Corp.
Event
Event 69596
Status
Terminated
Classification
Class II
Quantity
12 units
Official record key
device-enforcement:Z-0197-2015

Official wording

Reason: A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.

Code information: Lot Numbers PM0278 and PM0277

Distribution pattern: Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.