Skip to content
Recall Observatory FDA recall evidence

Device product

Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430. Stem Bodies Rejuvenate Modular consists of 6 modular stems bodies in sizes 7-12. The Modular Stem bodies are made of TMZF alloy, developed specifically for orthopaedic applications. Rejuvenate Modular Necks The Rejuvenate Modular Necks are made of Cobalt Chrome. The necks in this system consist of 4 lengths in 4 mm increments. They are 30 mm, 34 mm, 38 mm and 42 mm. They come in multiple neck angles and versions in each length. They are 127, 130 and 132. The versions are 0, 8, 16 degrees. In total there are 16 unique neck implants in the Rejuvenate Modular System.

Z-2090-2012

April 30, 2012

Class II

Product summary

Firm
Stryker Howmedica Osteonics Corp.
Event
Event 62456
Status
Ongoing
Classification
Class II
Quantity
(NLS/NLV) - 23193 units and (SPT) - 20,743 units
Official record key
device-enforcement:Z-2090-2012

Official wording

Reason: Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an

Code information: 510 K092561 all lots Rejuvenate (NLS/NLV) Rejuvenate (SPT) (4845-4XXX and NLS/NLV)

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an