Skip to content
Recall Observatory FDA recall evidence

Device product

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS, REF T653NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Z-0836-2014

September 17, 2013

Class II

Product summary

Firm
DeRoyal Industries Inc
Event
Event 66742
Status
Terminated
Classification
Class II
Quantity
409 units
Official record key
device-enforcement:Z-0836-2014

Official wording

Reason: The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Code information: Lot Numbers: 32901145, 33063999, 3132641, 3217556, 33276468, 33369411, 33412431, 33415413

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.