Skip to content
Recall Observatory FDA recall evidence

Device product

St Jude Medical, 10F Fast-Cath Trio" Hemostasis Introducer, REF # 406308, Sterile EO. The Fast-Cath Trio" Hemostasis Introducer consists of the following components: Port hemostasis adapter (3), Sheath (10F Cath-Lock"), dilator (10F) and 0.038 guidewire. See product labeling for diagram of contents.

Z-0108-2015

October 14, 2014

Class II

Product summary

Firm
St Jude Medical
Event
Event 69499
Status
Terminated
Classification
Class II
Quantity
220
Official record key
device-enforcement:Z-0108-2015

Official wording

Reason: St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.

Code information: Batch 4641580

Distribution pattern: Nationwide Distribution including FL, IN, ME, MA, NY, NY, OH, RI,TX, and WV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.