Device product
BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.035", 150 cm, REF SK15035M. Intended to support a guidewire during access into the vasculature.
Z-1802-2018
Product summary
- Event
- Event 79896
- Status
- Terminated
- Classification
- Class II
- Quantity
- 140 units
- Official record key
device-enforcement:Z-1802-2018
Official wording
Reason: Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present on the packaging hoops it may have the potential for cross-contamination in a clinical setting.
Code information: UDI: (01)00801741067921(17)191115(10)VTBY0563, Lot Number: VTBY0563
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present on the packaging hoops it may have the potential for cross-contamination in a clinical setting.