Skip to content
Recall Observatory FDA recall evidence

Device product

Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument.

Z-0849-2015

November 13, 2014

Class II

Product summary

Firm
Illumina Inc
Event
Event 69766
Status
Terminated
Classification
Class II
Quantity
13 units
Official record key
device-enforcement:Z-0849-2015

Official wording

Reason: It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer.

Code information: Serial No. M70130, M70189, M70108, M70165, M70166, M70175, M70172, M70103.

Distribution pattern: US Distribution to the states of AZ, GA, CA, OH, and VA., and one Internationally to France.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0