Skip to content
Recall Observatory FDA recall evidence

Device product

Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity in human serum.

Z-0933-2013

December 20, 2012

Class III

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 64063
Status
Terminated
Classification
Class III
Quantity
19, 503 cartons
Official record key
device-enforcement:Z-0933-2013

Official wording

Reason: The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).

Code information: Siemens Material Number (SMN) 10444944, Catalog number DF84 -- lot numbers BA3016, EXP 2013-01-16; EC3045, EXP 2013-02-14; EA3073, EXP 2013-03-14; DC3102, EXP 2013-04-12; FC3122, EXP 2013-05-02; EA3157, EXP 2013-06-06; BB3184, EXP 2013-07-03; EC3219, EXP 2013-08-07; EC3220, EXP 2013-08-08; GA3248, 2013-09-05; and FB3290, EXP 2013-10-17

Distribution pattern: Nationwide Distribution