Device product
Tibila Insert UC, Balanced Knee System, REF163-3YYY Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component
Z-2256-2014
Product summary
- Event
- Event 68898
- Status
- Terminated
- Classification
- Class II
- Quantity
- 13
- Official record key
device-enforcement:Z-2256-2014
Official wording
Reason: Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
Code information: UC Tibial Insert Sz 3 10mm,Part 163-3310,Lot A136109,UC Tibial Insert Sz 4 8mm,Part 163-3408,Lot A135948,UC Tibial Insert Sz 4 14mm,Part 163-3414,Lot A136110,UC Tibial Insert Sz 6 7mm,Part 163-3607,Lot A136113,UC Tibial Insert Sz 6 7mm,Part 163-3607,Lot A136746,
Distribution pattern: Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
Derived failure modes
-
Unknown
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.