Skip to content
Recall Observatory FDA recall evidence

Device product

Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable component of the Life Pulse High Frequency Ventilator. Model #(s): Individual Patient Circuit - catalog # 902 Patient Circuit Kit - catalog # 937: contains 2 circuits and 2 x 2.5 & 2 x 3.5 LifePort Adapters. Labeling states manufactured by: Bunnell Incorporated 436 Lawndale Drive, Salt Lake City, Utah 84115. Used for ventilating critically ill infants with pulmonary interstitial emphysema and infants with respiratory distress syndrome complicated by pulmonary air leaks, who are, in the opinion of their physicians, failing on conventional ventilation. The Patient Circuit is to provide a conduit for the humidification, warming, and temperature monitoring of the pressurized gas. The Patient Circuit is indicated for seven day single use.

Z-0514-2013

November 19, 2012

Class I

Product summary

Firm
Bunnell, Inc.
Event
Event 63710
Status
Terminated
Classification
Class I
Quantity
5743 patient circuits distributed in commerce
Official record key
device-enforcement:Z-0514-2013

Official wording

Reason: Customer complaints received indicate the heater wire insulation can melt, causing a short which creates sparking and smoke, in the circuit close to the humidifier cartridge. Bunnell is recalling all lots of circuits distributed between March 12, 2012 and November 30, 2012.

Code information: Patient Circuit (catalog # 902) lot # on each individual patient circuit label: 12C092, 12C102, 12C115, 12C125, 12C136, 12D159, 12D172, 12D189, 12E204, 12E211, 12E224, 12F241, 12F254, 12F271, 12G279, 12G290, 12G307, 12G321, 12H330, 12H349, 12I362, 12I371, 12I397, 12J413, 12J430, 12J448, 12K457, 12K471 Patient Circuit Kit (catalog # 397) lot # on the label on outside of box: 12C099, 12C114, 12C124, 12C135, 12D152, 12D170, 12D188, 12E203, 12E210, 12E223, 12F238, 12F255, 12F274, 12G209, 12G280, 12H334, 12H341, 12I356, 12I369, 12I392, 12J411, 12J425, 12J447, 12K467, 12K479

Distribution pattern: Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, Malaysia, and Uruguay (animal study only).

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Customer complaints received indicate the heater wire insulation can melt, causing a short which creates sparking and smoke, in the circuit close to the humidifier cartridge. Bunnell is recalling all lots of circuits distributed between March 12, 2012 and November 30, 2012.