Device product
Radiological Image Processing System The system is an aid to locate anatomical structures during open or percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgical technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CS, MR, or X-ray.
Z-2136-2012
Product summary
- Event
- Event 62414
- Status
- Terminated
- Classification
- Class I
- Quantity
- 17
- Official record key
device-enforcement:Z-2136-2012
Official wording
Reason: The FluoroTrak Spinal Navigation Application on the OEC 9900 EliteNAV could result in an incorrect position of the navigated instrument(s) versus the displayed reference image.
Code information: E9-0002-G, ES-0013-G, E9-0036-G, E9-0040-G, E9-0041-G, ES-0014-G, ES-0016-G, ES-0033-G, ES-0043-G, ES-0041-G, ES-0042-G, ES-OO84-G, ES-0086-G, ES-0163-G, E9-0089-G, ES-0026-G, ES-0044-G
Distribution pattern: Worldwide Distribution - USA including MO, IL, LA, FL, NH, TX, IN, PA, NJ and the countries of France and China
Derived failure modes
-
Unknown
The FluoroTrak Spinal Navigation Application on the OEC 9900 EliteNAV could result in an incorrect position of the navigated instrument(s) versus the displayed reference image.