Device product

ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems. For in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems.

Z-0340-2016

September 24, 2015

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 72392
Status
Terminated
Classification
Class II
Quantity
3031
Official record key
device-enforcement:Z-0340-2016

Official wording

Reason: Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.

Code information: 10379673 (160 tests kit), 10485591 (800 tests kit) Lot 230 (SMN 10485591 and 10379673) and lot 231(SMN 10485591)

Distribution pattern: Distributed in the states of AZ, NY, CA, IN, LA, MN, OH, MI, NM, and GA, and the countries of Mexico and Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.