ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems. For in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems.
Z-0340-2016
Product summary
- Event
- Event 72392
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3031
- Official record key
device-enforcement:Z-0340-2016
Official wording
Reason: Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.
Code information: 10379673 (160 tests kit), 10485591 (800 tests kit) Lot 230 (SMN 10485591 and 10379673) and lot 231(SMN 10485591)
Distribution pattern: Distributed in the states of AZ, NY, CA, IN, LA, MN, OH, MI, NM, and GA, and the countries of Mexico and Canada.
Derived failure modes
-
Unknown
Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.