siemens Dimension¿ Phenytoin Flex¿ Reagent Cartridge The PTN method used on the Dimension¿ clinical chemistry system is an in vitro diagnostic test intended to measure phenytoin (dilantin, diphenylhydantoin), an anti-epileptic drug, in human serum and plasma. PTN test results are used in the diagnosis and treatment of phenytoin overdose and in monitoring levels of phenytoin to ensure appropriate therapy.

Z-0148-2015

September 24, 2014

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 69372
Status
Terminated
Classification
Class II
Quantity
1733
Official record key
device-enforcement:Z-0148-2015

Official wording

Reason: Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with this lot resulting in the potential for a positive or negative bias and imprecision across the PTN assay range.

Code information: lot FA5154, exp. 2015-06-03

Distribution pattern: Worldwide Distribution - USA (nationwide) including Puerto Rico., and the countries of Australia, Brazil, Bahamas, Canada, Chile, India, Japan and Uruguay

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with this lot resulting in the potential for a positive or negative bias and imprecision across the PTN assay range.