Philips Healthcare Ingenuity CT X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Z-0017-2018

January 18, 2017

Class II

Product summary

Firm
Philips Electronics North America Corporation
Event
Event 78144
Status
Terminated
Classification
Class II
Quantity
US - 606
Official record key
device-enforcement:Z-0017-2018

Official wording

Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.

Code information: Software version 4.16

Distribution pattern: USA (nationwide) Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.