Device product

Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.

Z-2036-2012

June 12, 2012

Class II

Product summary

Firm
Teleflex Medical
Event
Event 62284
Status
Terminated
Classification
Class II
Quantity
14,794 each
Official record key
device-enforcement:Z-2036-2012

Official wording

Reason: Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.

Code information: Catalog #425-00, 425-10 and 425-30.

Distribution pattern: Worldwide Distribution - U.S. (nationwide) and the countries of Argentina, Australia, Brazil, Canada, Chile, China, Europe, India, Japan, Mexico, South Korea, Taiwan and UK.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.