The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.
Z-1849-2015
Product summary
- Event
- Event 71451
- Status
- Terminated
- Classification
- Class II
- Quantity
- 55
- Official record key
device-enforcement:Z-1849-2015
Official wording
Reason: After an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).
Code information: JAQ
Distribution pattern: Worldwide Distribution.
Derived failure modes
-
Unknown
After an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).