IMPAX CV Reporting The Results Management (RM)/IMPAX CV reporting component facilitates the quick, effective creating of digital structured reports for : adult catheterization and echocardiography, pediatric echocardiography, congenital heat disease, nuclear cardiology and non-invasive vascular disease management. This webb-based application allows cardiology departments to view, edit and sign the structured reports.

Z-2069-2012

July 10, 2012

Class II

Product summary

Firm
AGFA Corp.
Event
Event 62534
Status
Terminated
Classification
Class II
Quantity
6
Official record key
device-enforcement:Z-2069-2012

Official wording

Reason: When users selected "Left stenosis" in the "Graft Duplex Conclusion" section in IMPAX CV Reporting Non-Invasive(NIV) module, "Right stenosis" appeared in the sentence display within the printed representation of the report.

Code information: Software Version: RM 2.03 to 2.08, NIV 1.0

Distribution pattern: Nationwide Distribution including IN, SC, OH, TX

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0