Device product
Aplio 500/400/300 Diagnostic Ultrasound System (TUS-500/400/300). Device Listing Number: Dl53118. Diagnostic Ultrasound system.
Z-1104-2013
Product summary
- Event
- Event 64693
- Status
- Terminated
- Classification
- Class II
- Quantity
- 233 units
- Official record key
device-enforcement:Z-1104-2013
Official wording
Reason: Toshiba Medical Systems has received reports of startup issues with the Toshiba TUS-500 and TUS-300 ultrasound systems. The startup issue is predominant when attempting to bring the system up from standby mode when the system is used remotely. It has been determined that this may potentially cause damage to the system.
Code information: Item numbers: TUS-A300/Wl and TUS-A500/Wl
Distribution pattern: Nationwide Distribution.
Derived failure modes
-
Unknown
Toshiba Medical Systems has received reports of startup issues with the Toshiba TUS-500 and TUS-300 ultrasound systems. The startup issue is predominant when attempting to bring the system up from standby mode when the system is used remotely. It has been determined that this may potentially cause damage to the system.