Device product
REF 011-H3597 42 cm (16.5 IN) APPX 2.6ml, PUR TRANSFER SET, Clave ADDITIVE PORT, BCV, Spiros - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Z-1583-2021
Product summary
- Event
- Event 87703
- Status
- Ongoing
- Classification
- Class II
- Quantity
- Unavailable
- Official record key
device-enforcement:Z-1583-2021
Official wording
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Code information: REF: 011-H3597 UDI: (01) 1 0840619 06655 1 (17) 260101 (30) 50 (10) 5125225 Lot Number: 5125225
Distribution pattern: Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA and WA. The countries of Belgium, Canada, France, Italy, South Africa, Spain, United Kingdom.
Derived failure modes
-
Unknown
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.