Device product
Integrity 1.1. Integrity is the interface and control software for the Elekta range of medical digital linear accelerators and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes.
Z-1322-2014
Product summary
- Firm
- Elekta, Inc.
- Event
- Event 67574
- Status
- Terminated
- Classification
- Class II
- Quantity
- 498
- Official record key
device-enforcement:Z-1322-2014
Official wording
Reason: The problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values.
Code information: 152718, 152003, 152007, 151963, 152081, 151693, 152131, 152318, 105792, 105570, 152778, 151052, 152363, 152472, 152195, 153022, 152689, 152640, 152642, 152663, 152680, 152782, 152796, 152829, 152862, 152962, 153030, 153125, 153244, 153270, 153211, 153296, 152834, 152596, 151354, 151353, 152664, 151200, 151823, 151822, 151610, 152026, 152027, 152692, 151276, 152856, 151106, 152877, 152986, 152810, 105935, 153105, 153069, 151023, 152914, 151975, 151240, 151239, 153135, 105810, 153145, 153129, 151182, 151650, 152886, 151850, 152831, 153132, 151343, 153063, 105985, 152832, 152845, 152837, 152863, 152859, 152776, 152857, 152866, 152876, 152820, 152870, 152884, 152865, 152872, 152909, 152910, 152911, 152918, 152934, 152961, 152966, 153009, 153041, 153026, 153078, 153079, 153083, 153070, 153066, 153027, 153154, 153157, 153159, 153165, 153131, 153352, 152675, 152737, 152299, 152403, 152676, 151670, 152252, 151153, 151482, 151296, 152068, 152286, 152097, 151554, 151768, 151853, 151854, 151786, 151495, 151787, 153099, 152935, 151865, 151466, 105920, 152521, 151536, 151263, 105988, 151028, 151058, 152719, 151174, 151597, 152645, 152650, 152647, 105755, 152743, 151857, 151921, 105896, 151986, 152855, 152092, 151006, 151277, 105314, 105919, 152746, 152770, 105901, 151635, 152575, 152627, 151701, 151922, 151852, 152422, 152309, 152498, 152628, 152706, 152616, 105850, 152014, 151688, 151638, 152721, 153234, 152816, 105554, 105987, 151452, 151166, 152913, 151740, 152042, 152904, 152254, 152753, 152554, 152591, 152644, 152607, 152648, 152687, 152714, 152899, 152739, 152672, 152900, 152952, 152977, 153080, 153169, 153183, 152908, 153349, 153325, 152702, 152869, 152724, 151529, 152840, 152841, 151442, 151445, 151612, 151613, 151756, 105677, 151755, 151189, 151034, 151652, 152757, 105893, 105892, 152945, 151060, 105711, 105997. 151324, 151316, 151318, 151385, 152873, 151660, 152734, 152646, 152682, 151087, 151088, 105833, 152613, 152958, 105946, 151117, 151308, 151108, 151280, 151912, 152106, 152157, 152294, 152637, 152624, 152667, 152727, 152726, 152750, 152893, 152953, 151132, 151150, 151879, 151583, 152595, 152733, 153087, 152975, 153310, 153229, 152738, 152735, 152785, 152797, 152947, 153327, 152907, 153291, 153290, 152677, 152674, 151112, 105937, 151349, 105838, 105758, 105927, 151045, 151314, 151553, 151140, 151101, 105995, 105915, 151405, 152638, 152325, 152457, 152512, 151196, 151368, 152118, 152119, 105968, 151040, 105912, 151247, 105943, 151005, 105948, 151341, 151898, 151703, 151982, 152161, 151718, 153059, 151297, 151278, 151020, 105928, 152603, 153084, 152942, 151829, 151284, 152298, 151449, 151144, 151602, 152708, 152713, 152605, 152606, 152614, 152611, 152629, 152679, 152799, 152880, 152928, 152920, 152336, 153343, 153348, 152101, 151669, 151237, 151121, 153205, 151897, 105415, 152793, 105640, 105703, 105869, 151175, 151066, 151233, 152060, 151503, 152190, 153333, 152795, 151089, 152660, 152538, 152827, 152848, 152442, 152846, 152723, 152921, 151119, 152769, 152968, 152143, 152720, 151497, 151197, 151420, 105721, 151079, 105787, 105685, 151511, 151022, 105743, 105742, 151249, 151042, 105786, 151516, 151109, 151331, 151524, 152592, 151255, 105813, 151556, 151526, 105811, 151032, 151581, 151243, 151492, 152070, 151096, 151004, 151136, 105797, 105802, 105957, 105789, 151292, 151425, 151291, 152148, 151332, 151333, 151486, 151415, 151573, 151528, 151487, 151527, 151572, 151414, 105874, 151225, 152653, 105814, 151092, 151047, 151706, 151691, 151692, 151934, 151945, 151946, 153190, 151127, 105412, 152030, 152168, 152396, 152654, 151749, 152891, 152878, 152858, 105894, 151033, 151258, 151168, 151159, 151228, 152760, 151791, 152522, 151391, 151216, 151539, 151611, 151696, 151867, 152965, 151325, 152220, 151785, 152217, 151388, 151394, 151892, 152043, 151793, 151684, 152223, 152303, 152358, 151683, 152652, 152659, 152703, 151672, 152767, 152777, 152783, 152784, 152787, 152798, 152881, 151710, 153045, 152929, 153137.
Distribution pattern: Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CT, FL, GA, KY, LA, MD, MO, NM, NY, NC, OH, OK, PA, TN, TX, and WA and the countries of Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Poland, Romania, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
Derived failure modes
-
Unknown
The problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values.