Skip to content
Recall Observatory FDA recall evidence

Device product

TANGO3 Water Storage Tank with Ozone Disinfection System, Tango3, LLC, 132 Citizens Blvd., Simpsonville, KY 40067. Models are WOZ-250-PE, WOZ-240-SS, WOZ-180-SS. The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility.

Z-2138-2012

March 26, 2012

Class II

Product summary

Firm
Tango3 LLC
Event
Event 62514
Status
Terminated
Classification
Class II
Quantity
44 units
Official record key
device-enforcement:Z-2138-2012

Official wording

Reason: During an FDA inspection, the firm was notified of Quality System deficiencies, which included lack of procedures for design control and customized modifications that were either not reviewed or received proper clearance from FDA. There was no assurance that all systems in the field were working properly and operating at a minimum of 0.2ppm of ozone.

Code information: Model WOZ -250-PE, Serial #s: 1031, 1032, 1033, 1035, 1036, 1038, 1039, 1042, 1043, 1044, 1045, 1046, 1047, 1048, 1049, 1050, 1051, 1053, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1063, 1065, 1066, 1067, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076 & 1077. Model WOZ-180-SS, Serial #s: 1034, 1037, 1040 & 1064. Model WOZ-240-SS, Serial #s: 1041 & 1052.

Distribution pattern: Product was shipped to the following states: CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, ME, NH, NJ, NM, OH, PA, RI, TX and VA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During an FDA inspection, the firm was notified of Quality System deficiencies, which included lack of procedures for design control and customized modifications that were either not reviewed or received proper clearance from FDA. There was no assurance that all systems in the field were working properly and operating at a minimum of 0.2ppm of ozone.