Skip to content
Recall Observatory FDA recall evidence

Device product

AF w/AS 550 x 1-14ML/HR 2ML-15LO CPNB, Product code MVBT15XL-CPNB, Arrow AutoFuser Disposable Pain Control Pump

Z-0954-2019

October 02, 2018

Class II

Product summary

Firm
Arrow International Inc
Event
Event 81375
Status
Terminated
Classification
Class II
Quantity
19
Official record key
device-enforcement:Z-0954-2019

Official wording

Reason: Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Code information: 74B1702339 74H1601294 74D1701102 74H1700162 74D1702253 74H1701109 74E1701025 74H1701794 74F1700780 74J1700192 74F1701946 74J1700678 74G1600394 74J1701748 74G1700546 74K1702039 74G1701177 74L1700931 74G1702054

Distribution pattern: Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.