Device product
AF w/AS 550 x 1-14ML/HR 2ML-15LO CPNB, Product code MVBT15XL-CPNB, Arrow AutoFuser Disposable Pain Control Pump
Z-0954-2019
Product summary
- Event
- Event 81375
- Status
- Terminated
- Classification
- Class II
- Quantity
- 19
- Official record key
device-enforcement:Z-0954-2019
Official wording
Reason: Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Code information: 74B1702339 74H1601294 74D1701102 74H1700162 74D1702253 74H1701109 74E1701025 74H1701794 74F1700780 74J1700192 74F1701946 74J1700678 74G1600394 74J1701748 74G1700546 74K1702039 74G1701177 74L1700931 74G1702054
Distribution pattern: Nationwide distribution.
Derived failure modes
-
Unknown
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.