AF 550 x 5ML/HR CPNB, Product code MC0050XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Z-0949-2019
Product summary
- Event
- Event 81375
- Status
- Terminated
- Classification
- Class II
- Quantity
- 9
- Official record key
device-enforcement:Z-0949-2019
Official wording
Reason: Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Code information: 74C1601434 74E1602096 74C1602395 74F1602444 74C1703028 74F1700186 74D1700428 74G1702048 74D1701032
Distribution pattern: Nationwide distribution.
Derived failure modes
-
Unknown
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.