Device product
Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.
Z-1675-2015
Product summary
- Event
- Event 70959
- Status
- Terminated
- Classification
- Class II
- Quantity
- 4 units
- Official record key
device-enforcement:Z-1675-2015
Official wording
Reason: Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).
Code information: Part No: 0998-00-0708-01 Software version 8.9
Distribution pattern: Worldwide Distribution - US Nationwide in the state of TX.
Derived failure modes
-
Device software or design
Software anomaly