Device product
da Vinci Xi Surgical System, IS4000; ASSY,DISTAL SUJ,OUTER,IS4000, Material number: 380663-20 and 380663-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.
Z-1303-2015
Product summary
- Event
- Event 70705
- Status
- Terminated
- Classification
- Class II
- Quantity
- 167 total affected da Vinci XI systems
- Official record key
device-enforcement:Z-1303-2015
Official wording
Reason: Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.
Code information: Material number: 380663-20; Serial numbers: 1311140532, 1311140035, 1311140475, 1311140424, 1311140196, 1311140146, 1311140252, 1311140010, 1311140289, 1311140578, 1311140435, 1311140349, 1311140411, 1311140389, 1311140156, 1311140198, 1311140452, 1311140186, 1311140554 1311140335, 1311140249. Material number 380663-21: Serial numbers: 1311140047, 1311140541, 1311140181, 1311140352 1311140023, 1311140165, 1311140159, 1311140126, 1311140571, 1311140397, 1311140130, 1311140318, 1311140180, 1311140151 1311140375, 1311140428, 1311140528, 1311140263, 1311140169 1311140443, 1311140537, 1311140544, 1311140419, 1311140006 1311140583, 1311140173, 1311140154.
Distribution pattern: Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, China, France, Germany, India, Israel, Italy, Norway, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Derived failure modes
-
Unknown
Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.