Skip to content
Recall Observatory FDA recall evidence

Device product

0F 550 x 8ML/HR 2ML 60LO CPNB, Product code MT8060XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

Z-0944-2019

October 02, 2018

Class II

Product summary

Firm
Arrow International Inc
Event
Event 81375
Status
Terminated
Classification
Class II
Quantity
5
Official record key
device-enforcement:Z-0944-2019

Official wording

Reason: Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Code information: 74A1801899 74H1602701 74B1700146 74L1700201 74B1700921

Distribution pattern: Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.