ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Physiological patient monitor. Used in a hospital environment.

Z-2586-2014

July 02, 2014

Class II

Product summary

Firm
Spacelabs Healthcare Inc
Event
Event 69125
Status
Terminated
Classification
Class II
Quantity
8 units
Official record key
device-enforcement:Z-2586-2014

Official wording

Reason: The transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions. This is an expansion of recall Z-0871-2014 because additional serial numbers were identified as being part of this recall action as part of the firm's retrospective analysis of all recalls since 2011.

Code information: Serial numbers: 6281-002500, 6281-002501, 6281-002659, 6281-002661, 6281-002662, 6281-002663, 6281-002664, and 6281-003056. Model 96281, with SpO2 Option C.

Distribution pattern: US Distribution to TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions. This is an expansion of recall Z-0871-2014 because additional serial numbers were identified as being part of this recall action as part of the firm's retrospective analysis of all recalls since 2011.