Device product
Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010
Z-1925-2017
Product summary
- Event
- Event 76991
- Status
- Terminated
- Classification
- Class I
- Quantity
- 49 units
- Official record key
device-enforcement:Z-1925-2017
Official wording
Reason: LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Code information: All serial numbers
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.