Skip to content
Recall Observatory FDA recall evidence

Device product

Orbit soft Infusion sets, 9 mm soft cannula, P10 (42") tubing; Single-Use Set Orbit Infusion Sets are sterile, non-pyrogenic, single-use devices intended to be used for subcutaneous delivery of medication , for example insulin, from an external infusion pump.

Z-0990-2014

December 20, 2013

Class II

Product summary

Firm
Perrigo Diabetes Care
Event
Event 67154
Status
Terminated
Classification
Class II
Quantity
7924 units
Official record key
device-enforcement:Z-0990-2014

Official wording

Reason: Perrigo was notified by our supplier, Ypsomed AG, that they are recalling these batches due to being manufactured with a non-conforming female luer-lock connector. The luer-lock connectors are incompatible with standard male luer-lock connectors for use in connection to compatible infusion pumps. Perrigo became was notified of a quality complaint related to this issue on November 20, 2013 from a

Code information: 400064, 400069, 400070, & 400072 UPC: 7613117009502

Distribution pattern: Nationwide Distribution including GA & TX

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Perrigo was notified by our supplier, Ypsomed AG, that they are recalling these batches due to being manufactured with a non-conforming female luer-lock connector. The luer-lock connectors are incompatible with standard male luer-lock connectors for use in connection to compatible infusion pumps. Perrigo became was notified of a quality complaint related to this issue on November 20, 2013 from a