Device product
3.5 X 10 CORT SCREW SELFT 3.5 X 12 CORT SCREW SELFT 3.5 X 14 CORT SCREW SELFT 3.5 X 16 CORT SCREW SELFT 3.5 X 18 CORT SCREW SELFT 3.5 X 20 CORT SCREW SELFT 3.5 X 22 CORT SCREW SELFT 3.5 X 24 CORT SCREW SELFT 3.5 X 26 CORT SCREW SELFT 3.5 X 28 CORT SCREW SELFT 3.5 X 30 CORT SCREW SELFT 3.5 X 32 CORT SCREW SELFT 3.5 X 34 CORT SCREW SELFT 3.5 X 36 CORT SCREW SELFT 3.5 X 38 CORT SCREW SELFT 3.5 X 40 CORT SCREW SELFT 3.5 X 45 CORT SCREW SELFT 3.5 X 50 CORT SCREW SELFT 3.5 X 55 CORT SCREW SELFT 3.5 X 60 CORT SCREW SELFT 3.5 X 75 CORT SCREW SELFT
Z-1044-2014
Product summary
- Firm
- Zimmer, Inc.
- Event
- Event 66500
- Status
- Terminated
- Classification
- Class II
- Quantity
- 7,044,680 total
- Official record key
device-enforcement:Z-1044-2014
Official wording
Reason: Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Code information: 61173204 61173205 61173206 61804309 61173207 61804310 61173208 61117130 61173210 61173212 61173213 61173214 61173215 61173216 61173218 61173219 61173220 61173221 61173223 61173224 61173226 61173228 61173231
Distribution pattern: Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Derived failure modes
-
Unknown
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.