Device product
Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the following CT systems: Aquilion ONE TSX-301C X-ray CT system with a 320-row 0.5 mm detector is provided with a 7.5-MHU large-capacity X-ray tube unit, and permits scanning over large areas at the high-speed of 0.35 s per rotation. The system features are; 160-mm-wide area detector, station, gantry and patient couch operation controls, image reconstruction, high-speed volume scan, high-quality images, selectable image slice thickness, dual-monitor system, improved image diagnosis functions, high-speed volume data workflow, exposure reduction, ECG-gated scan and reconstruction, and SUREFluoro. The CT cardiac function analysis software is used to analyze cardiac function using data obtained by performing cardiac MPR processing for ECG-gated scan data according to the cardiac axis. Aquilion ONE is a dynamic volume CT system that supports whole-body scanning. This 320-detector row system generates 640 slices per rotation using the coneXact reconstruction algorithm. In addition, the high-speed rotation mechanism and the fast reconstruction unit of the system allow the rapid image acquisition to further improve throughput in CT examinations.
Z-0471-2015
Product summary
- Event
- Event 69748
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1 unit within the U.S. only
- Official record key
device-enforcement:Z-0471-2015
Official wording
Reason: Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated based on some of those parameters.
Code information: Item CFA Software Serial Number: Description TSX-301C/2D 2BA1362093 Aquilion One Vision
Distribution pattern: US Nationwide Distribution to CA, MA, MD
Derived failure modes
-
Device software or design
software (CFA). Incorrect