Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.

Z-0118-2013

August 21, 2012

Class II

Product summary

Firm
Elekta, Inc.
Event
Event 63405
Status
Terminated
Classification
Class II
Quantity
52 units
Official record key
device-enforcement:Z-0118-2013

Official wording

Reason: During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.

Code information: None as the product is software

Distribution pattern: Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.