Device product

VistaSolution, physiological signal transmitter. A software graphical user interface for use by healthcare professionals to display physiological data collected by the wireless remote monitoring system in healthcare settings. This is a secondary, adjunct patient monitor and is not intended to replace existing standard-of-care patient monitoring.

Z-2537-2020

May 21, 2020

Class II

Product summary

Firm
Vitalconnect Inc.
Event
Event 85802
Status
Terminated
Classification
Class II
Quantity
7
Official record key
device-enforcement:Z-2537-2020

Official wording

Reason: A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification. The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.

Code information: Catalog number: 6200, Vista Solution version 2.1, UDI: 00850490007139

Distribution pattern: U.S. Nationwide distribution including in the states of NJ, NY, VA, OUS: None

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software defect