Skip to content
Recall Observatory FDA recall evidence

Device product

Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UDI): D7915147A10 - Product Usage: are intended for CAD/CAM fabrication of all-ceramic (no metal) dental restorations. Luminesse ZR Blanks are partially sintered Yttria stabilized zirconia blanks for use as CAD/CAM milling blanks.

Z-2943-2020

August 06, 2020

Class II

Product summary

Firm
Talladium Inc
Event
Event 86227
Status
Terminated
Classification
Class II
Quantity
39 pieces
Official record key
device-enforcement:Z-2943-2020

Official wording

Reason: The linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was outside the acceptable range. The acceptable average linear shrinkage ranges between 1.225 to 1.275. The production lot being recalled has a linear shrinkage rate of 2.251 which is above the acceptable range.

Code information: Lot Number.: 1AZG

Distribution pattern: U.S. Nationwide distribution including in the states of CA, FL, KS, LA, MO, and PA. No OUS Consignees.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    label is incorrect