Device product
Fresenius 2008T Series Hemodialysis System
Z-1026-2017
Product summary
- Event
- Event 76037
- Status
- Terminated
- Classification
- Class II
- Quantity
- 65361 units (US) 249 units (OUS)
- Official record key
device-enforcement:Z-1026-2017
Official wording
Reason: When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
Code information: All serial numbers
Distribution pattern: Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
Derived failure modes
-
Unknown
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine