Device product
Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
Z-1487-2020
Product summary
- Event
- Event 84756
- Status
- Terminated
- Classification
- Class II
- Quantity
- 70
- Official record key
device-enforcement:Z-1487-2020
Official wording
Reason: A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Code information: Serial Codes 3217017 200 58 72 CV.170 180 524615 47 211 364 30 3862570/0011 529 S.N. 519 S01H060871 505/0050114 S01H057274 01H077213 530 532 S01HA52566 348 476 320 481 467 284 352 359 459 73*004 733 536 CV.474 454 49 281 51 186 122 250 37 22 CV.312 296 373 CV.596 176 30 580 279
Distribution pattern: Nationwide domestic distribution. Foreign distribution worldwide.
Derived failure modes
-
Unknown
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.