ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954

Z-1140-2022

October 22, 2021

Class II

Product summary

Firm
MEDLINE INDUSTRIES, LP - Northfield
Event
Event 90126
Status
Terminated
Classification
Class II
Quantity
6 boxes
Official record key
device-enforcement:Z-1140-2022

Official wording

Reason: Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be provided due to probable Quality Control failure. A delay in diagnosis may be experienced with repeat testing resulting in patient (customer) inconvenience, pain, and cost of the repeat blood draw and testing. All HII1220001954 product shipped/stored at a Medline branch between July 2021 and August 2021 in the identified lots were incorrectly stored and shipped. Six (6) boxes have been shipped to end users at room temperature instead of refrigerated (ice). Five (5) boxes are on hold at a Medline warehouse.

Code information: Model Number: HII1220001954; Lot Numbers: 215907578 (Expiration Date 09/30/2022), 215907729 (Expiration Date 02/28/2023); Manufacturer Number: A11A01954

Distribution pattern: US Distribution to North Carolina.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    Quality Control failure