NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900
Z-1251-2022
Product summary
- Event
- Event 90275
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 459 kits US; 1626 kits OUS
- Official record key
device-enforcement:Z-1251-2022
Official wording
Reason: There is a potential for reduced analytical sensitivity for samples in the 1x - 3x LOD range for Flu A, RSV, and SARS-CoV-2 targets. This may lead to false negative results, allowing ongoing transmission to others and delaying appropriate treatment.
Code information: UDI-DI (GTIN): 10814278020830 Cartridge lot numbers 111915 - 117508
Distribution pattern: Domestic distribution to FL, NY, MD, PA, VA, MI, TN, NM. Foreign distribution to AT, AU, BE, CH, DE, FI, FR, GB, GR, HK, ID, IT, LT, NL, SA, SE, and SI.
Derived failure modes
-
Unknown
There is a potential for reduced analytical sensitivity for samples in the 1x - 3x LOD range for Flu A, RSV, and SARS-CoV-2 targets. This may lead to false negative results, allowing ongoing transmission to others and delaying appropriate treatment.