Skip to content
Recall Observatory FDA recall evidence

Device product

Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK

Z-0032-2022

August 09, 2021

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 88559
Status
Ongoing
Classification
Class II
Quantity
293,640 kits
Official record key
device-enforcement:Z-0032-2022

Official wording

Reason: Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples.

Code information: Catalog Number:B52699 Lot Number / UDI Code: 921491 / (01)15099590693183(17)201130(11)191201(10)921491; 921638 / (01)15099590693183(17)201231(11)200101(10)921638; 921631 / (01)15099590693183(17)210131(11)200201(10)921631; 921750 / (01)15099590693183(17)210131(11)200201(10)921750; 921805 / (01)15099590693183(17)210228(11)200229(10)921805; 921831 / (01)15099590693183(17)210331(11)200331(10)921831; 921943 / (01)15099590693183(17)210430(11)200430(10)921943; 922000 / (01)15099590693183(17)210531(11)200531(10)922000; 922039 / (01)15099590693183(17)210630(11)200630(10)922039; 922040 / (01)15099590693183(17)210731(11)200731(10)922040; 922062 / (01)15099590693183(17)210731(11)200731(10)922062; 922145 / (01)15099590693183(17)210831(11)200831(10)922145; 922146 / (01)15099590693183(17)210930(11)200930(10)922146; 922338 / (01)15099590693183(17)210930(11)200930(10)922338; 922440 / (01)15099590693183(17)211031(11)201031(10)922440; 922545 / (01)15099590693183(17)211031(11)201031(10)922545; 922547 / (01)15099590693183(17)211130(11)201130(10)922547; 922713 / (01)15099590693183(17)211231(11)201231(10)922713; 922714 / (01)15099590693183(17)211231(11)201231(10)922714; 922722 / (01)15099590693183(17)220131(11)210131(10)922722; 922897 / (01)15099590693183(17)220131(11)210131(10)922897; 124235 / (01)15099590693183(17)220228(11)210228(10)124235; 124339 / (01)15099590693183(17)220228(11)210228(10)124339; 922898 / (01)15099590693183(17)220228(11)210228(10)922898; 124340 / (01)15099590693183(17)220331(11)210331(10)124340; 124348 / (01)15099590693183(17)220430(11)210430(10)124348; 124353 / (01)15099590693183(17)220430(11)210430(10)124353; 124688 / (01)15099590693183(17)220430(11)210430(10)124688; 124354 / (01)15099590693183(17)220531(11)210531(10)124354; 124355 / (01)15099590693183(17)220630(11)210630(10)124355; 124356 / (01)15099590693183(17)220630(11)210630(10)124356; 124357 / (01)15099590693183(17)220630(11)210630(10)124357;

Distribution pattern: U.S.: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, NC, ND, NE, NV, NJ, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY O.U.S.: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, China, Croatia, Cura¿ao, Czech Republic, Denmark, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Niger, Nigeria, Norway, Oman, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Vietnam, Yemen, and Zimbabwe

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples.